Summary
LZZ–S Double Rotor Flowmeter is high precision measuring instruments used for continuous measurement of liquid flow in pipeline. It has no pulse, low noise, high precision, high reliability, wide range, small pressure loss, strong adaptability of viscosity, and can measure the high temperature, high viscosity liquids, calibration of convenient, easy to install, and many other advantages.
The meter is equipped with the pointer display, word wheel cumulative counting device, and can directly display cumulative flow when liquid flows through the pipe, it can also realize remote display through connection between the transmitter and the electronic instrument or computer. The flowmeter is suitable for petrochemical, chemical, chemical fiber, traffic, commerce, food, medicine, health care and other industrial departments of commerce measurement control and engineering management.
  • Operating Principle
  • Technical Parameter
  • Model Selection Table
  • Outline and installation
  • Ordering information
LZZ-S double rotor flowmeter consists of flow transmitter and counter. The main part of the transmitter is composed of measuring room equipped with a pair of special tooth profile of helical rotor and seal coupling. Counting mechanism contains retarding mechanism, precision adjusting mechanism, counter for print wheel accumulation and electrical transmitter, etc.
Measuring room mainly composed of a pair of special rotor and cover seal cavity as flow measurement units. Rotor powered by the differential pressure of import and export of flowmeter so that send the inlet liquid to the exit after passing the sealing cavity measurement. The amount of liquid through seal chamber every turn is eight times of the picture, sealing coupling transfers the total rotation of the rotor and rotating speed to the counting mechanism, and a pointer to display and print wheel accumulation, then you can know the total amount of liquid passing by the pipe. You can also install transmitter in counting mechanism, namely become the flowmeter which can output signal, the transmitter matches with our company’s electric display instrument, can also realize remote transmission (quantitative, accumulated, moments, etc) automatic measurement and control. You can see the instruction in the display instrument classification.


1. counter  2. transmitter 3. precision adjusting mechanism 
4. flow transmitter 5.external precision adjustment device
 
1 main parameters
1.1 Accuracy class:0.2%;0.5%(Note:If you have special requirements we can provide the accuracy class of 0.1%)
1.2 Nominal diameter: 25;40;50;80;100;150;200;250;300mm;
1.3 Pressure class: 1.6MPa;2.5MPa;4.0MPa;6.3MPa;
1.4 Flange standard: HG/T20592-2009,HG/T20617-2009 or on request.
1.5 Fluid viscosity: 0.3-3500mPa.s(dynamic viscosity)
1.6 Fluid temperature: common temperature-25℃~+100℃;high tempreature+100℃~+250℃(Please indicate the special temperature )
1.7 Flow range as shown in the table below

(Unit: m³/h; Accuracy class: 0.5)

 
Model diameter viscosity mPa.S
0.3~0.8 0.8~2 2~15 15~400 400~1000 1000~2000 2000~3500
gasoline kerosene diesel
engine oil
heavy oil high viscosity liquid
LZZ-S-25B 25 3~9 1.5~10 1~10 1~10 1~8 1~8 1~6
LZZ-S-40A 40 3~9 1.5~10 1~10 1~10 1~8 1~8 1~6
LZZ-S-40B 7~20 3~22 2.5~25 2.5~22 2~18 2~18 3~12
LZZ-S-50A 50 7~20 3~22 2.5~25 2.5~22 2~18 2~18 3~12
LZZ-S-50B 9~36 4.5~36 3.6~36 3.6~36 2.8~25 2.8~25 4.5~18
LZZ-S-80A   9~36 4.5~36 3.6~36 3.6~36 2.8~25 2.8~25 4.5~18
LZZ-S-80B 80 20~80 10~80 10~100 10~100 6.5~56 6.5~56 5~40
LZZ-S-100A 100 20~80 10~80 10~100 10~100 6.5~56 6.5~56 5~40
LZZ-S 型-100B 25~100 20~100 15~150 15~150 8.5~80 8.5~80 6.5~55
LZZ-S-150A 150 25~100 20~100 15~150 15~150 8.5~80 8.5~80 6.5~55
LZZ-S-150B 55~225 30~250 25~250 25~250 18~150 18~150 10~100
LZZ-S-200A 200 55~225 30~250 25~250 25~250 18~150 18~150 10~100
LZZ-S-200B 90~360 50~400 40~400 40~400 28~240 28~240 20~160
LZZ-S-250A 250 90~360 50~400 40~400 40~400 28~240 28~240 20~160
LZZ-S-250B 130~540 65~540 60~600 60~600 42~360 42~360 30~240
LZZ-S-300A 300 130~540 65~540 60~600 60~600 42~360 42~360 30~240
LZZ-S-300B 220~800 110~900 95~950 95~950 70~600 70~600 54~450

(Unit: m³/h; Accuracy class:0.2)
 
Model diameter viscosity mPa.S
0.3~0.8 0.8~2 2~15 15~400 400~1000 1000~2000 2000~3500
gasoline kerosene diesel
engine oil
heavy oil high viscosity liquid
LZZ-S-25B 25 3.5~8 3~10 2~10 2~10 2~8 2~8 2~6
LZZ-S-40A 40 3.5~8 3~10 2~10 2~10 2~8 2~8 2~6
LZZ-S-40B 8~20 5.5~22 4.5~22 4.5~22 4~18 4~18 3~12
LZZ-S-50A 50 8~20 5.5~22 4.5~22 4.5~22 4~18 4~18 3~12
LZZ-S-50B 15~36 9~36 7~36 7~36 6~25 6~25 4.5~18
LZZ-S-80A 80 15~36 9~36 7~36 7~36 6~25 6~25 4.5~18
LZZ-S-80B 30~80 20~80 15~80 15~80 14~56 14~56 10~40
LZZ-S-100A 100 30~80 20~80 15~80 15~80 14~56 14~56 10~40
LZZ-S-100B 40~100 25~100 20~120 20~120 18~72 18~70 14~55
LZZ-S-150A 150 40~100 25~100 20~120 20~120 18~72 18~70 14~55
LZZ-S-150B 88~220 58~225 44~220 44~220 38~150 38~150 25~100
LZZ-S-200A 200 88~220 58~225 44~220 44~220 38~150 38~150 25~100
LZZ-S-200B 150~360 90~360 72~360 72~360 50~210 50~210 40~160
LZZ-S-250A 250 150~360 90~360 72~360 72~360 50~210 50~210 40~160
LZZ-S-250B 180~540 135~540 100~540 100~540 90~360 90~360 60~240
LZZ-S-300A 300 180~540 135~540 100~540 100~540 90~360 90~360 60~240
LZZ-S-300B 250~800 220~900 180~900 180~900 150~600 150~600 110~450

2 The major component material and the nominal pressure

 
Model Housing,front and back cover covering plate Rotor Axis Axle sleeve Nominal pressure(MPa)
LZZ-S-C Carbon steel Carbon steel cast iron、aluminium alloy,Stainless Steel Stainless Steel Graphite Below DN100 is 4.0;Beyond DN150 is 2.5
LZZ-S-P Stainless Steel Stainless Steel Stainless Steel Graphite / ball bearing 1.6、2.0、2.5、
LZZ-S-S double housing Specification:DN80、DN100、DN150 Nominal pressure 6.3NPa;DN200 Nominal pressure 4.0MPa;

3 Double rotor flowmeter with warm jacket
In order to facilitate transporting and measuring flow for the medium that is easy to freeze under normal temperature or condense, crystallize, usually need to heat medium in the system pipe, and melt the medium and hold its temperature. But due to the double rotor flowmeter installation does not allow directly access to the steam (purpose is to prevent damage meter), therefore we design the flowmeter with warm jacket. The flowmeter housing add warm jacket outside, to melt solidified medium in the flowmeter, you can use hot water, hot oil or below 200 ℃ steam flux into the warm jacket and keep certain temperature so that make sure the flowmeter keep moving.

The filter installed on double rotor flowmeter can also be adopted with the structure of warm jacket.
The inlet and outlet of warm jacket usually is designed by DN15 diameter flange, or on special request. 
Model Code Content
LZZ-S   Double rotor flowmeter
  25   Nominal Diameter:DN25mm
40   Nominal Diameter:DN40mm
50   Nominal Diameter:DN50mm
80   Nominal Diameter:DN80mm
100   Nominal Diameter:DN100mm
150   Nominal Diameter:DN150mm
200   Nominal Diameter:DN200mm
250   Nominal Diameter:DN250mm
300   Nominal Diameter:DN300mm
  C   The structure of material:Carbon Steel
P   The structure of material:Stainless Steel
  1.6   Nominal Pressure PN1.6MPa
2.5   Nominal Pressure PN2.5MPa
4.0   Nominal Pressure PN4.0MPa
  6.3   Nominal Pressure PN6.3MPa
  JS 型 型 Counter:local indication;
  GF With transmitter,output signal:pulse;
  MF With transmitter,output signal:4~20mA;
    i Explosion protection type:Intrinsically safe ExiaIICT5
  d Explosion protection type:Explosion isolation ExdIICT6
  C Normal temperature
G High temperature
  D Single housing
S 型 型 Double housing
  B Warm jacket
1 The structure and dimension
1.1 The structure and dimension of the flowmeter


 
Model diameter L H B C I D D1 N d0 Pressure
MPa
Reference
weight
standard Double housing
LZZ-S-25B 25 300   420 175 228 90 115 85 4 14 The flange
example:
4.0MPa
29
LZZ-S-40A 40 300   420 175 228 90 150 110 4 18 33
LZZ-S-40B 40 300   450 200 258 112 150 110 4 18 40
LZZ-S-50A 50 300   450 200 258 112 165 125 4 18 42
LZZ-S-50B 50 340   525 240 333 150 165 125 4 18 62
LZZ-S-80A 80 340   525 240 333 150 200 160 8 18 64
LZZ-S-80B 80 380 400 580 285 388 175 200 160 8 18 94
LZZ-S-100A 100 380 400 580 285 388 175 235 190 8 22 98
LZZ-S-100B 100 440 450 660 339 468 250 235 190 8 22 146
LZZ-S-150A 150 440 450 660 339 468 250 300 250 8 26 152
LZZ-S-150B 150 500 560 740 410 548 270 300 250 8 26 238
LZZ-S-200A 200 500 560 740 410 548 270 360 310 12 26 245
LZZ-S-200B 200 550 700 820 455 628 285 360 310 12 26 The flange
example :
2.5MPa     337
337
LZZ-S-250A 250 550 700 820 455 628 285 425 370 12 30 350
LZZ-S-250B 250 700   910 550 718 350 425 370 12 30 570
LZZ-S-300A 300 700   910 550 718 350 485 430 16 30 600
LZZ-S-300B 300 800   1010 645 818 415 485 430 16 30 855

Note:  (1) "H"dimension prepare for G the dimension of the governor;
 (2) The flowmeter with warm jacket, the inlet and outlet of heat source is DN15 flange connection;
 (3) If flowmeter configures heat radiator, please add the heat radiator size to H size in the table, it is total height of high temperature flowmeter;
(4)Double housing flowmeter pressure can be up to 6.3 MPa; DN200 pressure is 4.0 MPa. Special designment can be up to 10 MPa;

1.2The radiator structure and dimension

Note: 
If the model of the flowmeter is G, the flowmeter is high temperature, it is used for 60-120℃, without transmitter and may not configure heat radiator. 
If the model of the flowmeter is G2, the flowmeter is high temperature, it is used for 120-250℃, and configure long heat radiator.
If the model of the flowmeter is G1, the flowmeter is high temperature, it is used for 60-120℃, and with transmitter, configuring short heat radiator.

2 Installation and application

2.1 The filter should be installed in front of the flowmeter, their housing’s arrow directions are the same.
2.2 When the measured liquid containing gas, the gas separator should be installed in front of flowmeter.
2.3 Under the condition of correct installation of the flow meter, if it is not easy to see reading, counter can be turned to 180 ° or 90 °.
2.4 Push the rotor a few times with a wooden stick from the exit before installing a new flowmeter, if the rotor does not move, you can soak with gasoline (to avoid deposits after factory check).
2.5 Throttle valve should be installed in the inlet of the flow meter, on-off valve installed in the exit, When using on-off valve needs a slow start, don't suddenly open valve.
2.6 It is forbidden to use steam sweep line through the flowmeter.
2.7 In continuous occasion, flowmeter must add by-pass pipe.
2.8 The pipe shall be flush before flowmeter installation, when flushing straight pipe (replace flow meter location) to prevent welding slag, debris into the meter.
2.9 It is forbidden to check meter composed of carbon steel material by water.
2.10 When flow meter measuring flow, the value of flow shall not exceed the range of flow range. Flow meter working in 40-80% of maximum flow is suitable.

3 Error calculation and adjustment
3.1 The basic error of flowmeter is calculated by the verification flow point of all measured values according to the following formula: (volume)
E=(Vm-V)/V×100%
E--The error of flowmeter(generally refers to accumulative error) take two significant figures.
Vm--Flowmeter measuring values(indicating value)
V--After meter is corrected, flow standard device measuring value ( the actual value)
From the basic formula of error, when Vm is greater than V, The basic error of flowmeter is“+”, it indicates flowmeter going fast.
From the basic formula of error, when Vm is less than V, The basic error of flowmeter is“-”, it indicates flowmeter going slow. 
In order to make the flowmeter errors within a basic error limits, often need to adjust the error. By replacing a pair of adjusting gear (tooth) in counter changes the mechanical transmission speed ratio, so that the flow meter’s indicating value gets adjusted.
Error adjustment can't change features of the flowmeter, only make its features curve artificially in a new coordinate. In general, in the prescribed (or actual use) flow range, the basic error range of the maximum and minimum flow check point is not more than the precision of the basic error limits, all can realize flowmeter basic error qualified by adjusting error. 
For flow meter has been used, usually use the original adjusting gear checking error, and then adjusting error according to the error of the specific situation.
3.2 Double rotor flowmeter error adjusting method (error adjustment instructions)
3.2.1 Common error adjusting method:
3.2.1.1 In designing the standard of the double gear is 38/35, During inspection, if you find the meter is going faster, namely appear error (+),such as +1.02-+0.3,the gear should be changed 38/35 into 41/38, at this time original point of the error curve moves up from 38/35 gear corresponding to zero to 41/38 gear to the location of + 0.63 (see table), so error curve in a new coordinate, and the error of flowmeter has been adjusted in the range of + 0.39 ~ ﹣0.33,and satisfy the qualified requirements(see figure 1).
3.2.1.2When the flowmeter is working, due to the different working conditions and changes, the error usually changes and it may exceed the maximium error, if only the error range is less than 1%, generally can be transferred to qualified, for example, instrument error drops to 0.7 ~ + 0.2, and need to change the double gear, at first you should know that how many gear teeth the double gear has. If the double gear is38/35, adjust according to the above of the first method. If the double gear is38/35, then take the error+ 0.63 corresponding to gear as zero(the origin of coordinates).Now the 41/38 gear should be changed into the double gear of 41/37, error can be adjusted to the range of﹣0.5-﹢0.4 that make sure qualified (see figure 2).

3.2.1.3 The specific method of calibrating and adjusting double rotor flowmeter of class 0.2 should be according to the national metrological verification procedures JJG667-1997 "liquid volumetric flowmeter verification regulation".
3.2.2 The error adjustment method of GF3 external precision regulator
GF3 is a kind of external precision regulator, when error adjusting, do not need to remove the governor and can be directly adjusted in the external, the operation is convenient, easy and accurate.
Example for adjustment method:
When there is + 0.63 ~ 0.67% error in calibration flowmeter, after unscrewing the lid of the external regulator, you can see the left and right knobs, adjust error according to the surface of the scale of the external precision regulator. Coarse adjustment specification is 0.45% per cell, adjusting the coarse gear one cell in the direction of negative, that is﹣0.45%,it can make the original error theory adjusted + 0.63 ~ + 0.67% to + 0.18 ~ + 0.22%, then you can adjust fine tuning gear two cell in the direction of negative. The value of adjustment is 0.05×2=0.10%, then the adjustment is from + 0.18 ~ + 0.22% to + 0.08 ~ + 0.12, achieved the best error value.
Suggest high accuracy (class 0.2) flowmeter configuring GF3 is preferred.
3.3 Error adjustment table 
 
Error adjustment quantity﹪ The set of adjusting the gear Error adjustment quantity﹪ The set of adjusting the gear
Above Z Below Z Above Z Below Z
4.21 33 30 0.28 40 38
3.90 34 31 0.40 41 39
3.62 35 32 0.53 42 40
3.35 36 33 0.64 43 41
3.10 37 34 0.75 44 42
2.86 38 35 0.86 45 43
2.63 39 36 0.96 46 44
2.42 40 37 1.14 24 23
2.22 41 38 1.32 25 24
2.02 42 39 1.47 26 25
1.84 43 40 1.62 27 26
1.75 29 27 1.75 28 27
1.67 44 41 1.88 29 28
1.50 30 28 2.00 30 29
1.35 46 43 2.11 31 30
1.27 31 29 2.21 32 31
1.05 32 30 2.30 33 32
0.85 33 31 2.39 34 33
0.66 34 32 2.48 35 34
0.48 35 33 2.63 37 38
0.31 36 34 2.77 39 38
0.15 37 35 2.89 41 40
0.00 38 36 3.01 43 42
0.14 39 37 3.16 46 45
When instrument value is less than the actual value,choosing above Z, below Z from down to up When instrument value is greater than the actual value, choosing above Z, below Z from down to up


4 Match with transmitter and accessories
Installing a transmitter on the double rotor flowmeter (or transducer)of our company, so it can be remote transmitting, it is transferred the flow of measured medium into pulse signal, Thus realize the automatic control system.
4.1 GF transmitter
GF transmitter is pulse transmitter device matching with double rotor flowmeter used in ExiaIICT5 and ExdIICT6 level places. Explosion-proof form is intrinsically safe and explosion isolation respectively, on the housing and nameplate have obvious explosion-proof symbol, after GF transmitter and safety barrier is connected, and they should match with display instrument.
GF-ⅠTransmitter technology characteristic
1、operating voltage DC12V±10%
2、output signal:Voltage pulse high level greater than 9 v;Voltage pulse low level less than 0.5 v
3、Match with safety barrier:KN9714C
4、three-wire system(VDD   signal  GND)
5、The intrinsically safe connection diagram
6、Common and explosion isolation interface circuit diagram
GF-ⅡTransmitter technology characteristic
1、operating voltage: DC24V±10%
2、output signal:Voltage pulse high level greater than 20 v;Voltage pulse low level less than 1 V 
3、Match with safety barrier:KN9714 
4、three-wire system(VDD  signal  GND)
5、The intrinsically safe connection diagram
6、Common and explosion isolation interface circuit diagram
 


4.2 MF series 4 ~ 20mA analog output transmitter
4.2.1 Summary
MF series transmitter can match with double-rotor flowmeter of our company, instantaneous flow is transferred into 4 ~ 20 mA analog output on scene, and remote to control room, for the instantaneous flow display, adjust and control. At the same time the output flow pulse signals used for calculating flow value.
4.2.2 Feature
4.2.2.1 Four-wire system(VDD line、4-20mA electric streamline、Pulse signal lines、GND line)
Note: Only to 4 ~ 20 mA output for three-wire system, pulse signal lines do not supply
4.2.2.2 Flowmeter directly outputs 4 ~ 20 mA, long transmission distance, safe and reliable.
4.2.2.3 Explanation of Part Numbers



4.2.3 Technology parameters
4.2.3.1Output signal: a、4 ~ 20 mA analog
                 b、Voltage pulse:VL<0.5V VH>9V 12V; VL<1V VH>20V 24V
4.2.3.2 Allow error  analog:±0.5%FS
                 pulse:±1pulse
4.2.3.3 Ambient temperature ﹣41~﹢60℃(Explosion-proof requirements﹣10~+60℃)
4.2.3.4 The analog load resistance (client)  <400Ω power supply12 V;<800Ω power supply24 V
4.2.3.5 Power supply:DC12V±10%60mA; DC24V±10%60mA
4.2.3.6 Explosion-proof sign: ExiaIICT6、ExdIICT6
4.2.4 The user instructions
When ordering, please indicate the power supply is 24 V DC or 12V DC. And wiring referring to the following figure



4.3 Safety barrier outline and installation dimensions (JF)
Note: the company can matching JF type safety barrier, user must indicate when ordering


4.4 Pulse transmitter and double rotor flowmeter match with transmitting parameter table
 
DN L/P P/S DN L/P P/S
Model A5+A25T+GF Model A5+G+GF
25B 0.01 250 150A 0.1/1 417/41.7
40A 0.01 250 150B 1 69.4
口径 A5+G+GF 200A 1 69.4
40B 0.1   200B 1 94.4
50A 0.1 69.4 250A 1 94.4
50B 0.1 111.11 250B 1 138.9
80A 0.1 111.11 300A 1 138.9
80B 0.1 277.8 300B 1 222.2
100A 0.1 277.8      
100B 0.1/1 417/41.7      

5 Fault analysis and elimination
 
fault phenomenon reason analysis elimination methods
The rotor do not run 1、There are sundries in pipe
2、Measured liquid contains many sundries .The filter is damaged. Sundries go into meter,the gear get stuck.
Unpick and flush instrument and pipeline, repair the filter
Axial seal coupling leakage Seal packing wear or lack of seal oil Tighten gland or replace filler, filling sealing oil.
The point rotates unsteadily or go intermittently Pointer, washer, etc loose or rotor rotating inflexible. Retighten, eliminate the inflexible phenomenon.
Small flow negative error is too large Rotor collide with metering box wall, the reason is bearing wear, or metering box wall deformation Replace bearings, repair metering box walls and gear of variable tooth, make rotation of the gear flexible, make sure the required clearance
Error change too much Fluid has big pulsating or containing gases. Reduce pulse or add gas separator
Error is too big, but the difference between the maximum and minimum error less than 1% Beyond usage period or after  repairing the clearance change over Check and adjust again
Transmitter has no signal Inappropriate position of transmitting piece To adjust the position, the left and right sides, back and forth motion
polarity reversed Back to connect the "+" red lines "-" the black line
◇Model and specification:       ◇Accuracy class:      ◇Maximum and minimum flow value:
◇Measured medium:           ◇Operating temperature:     ◇Medium viscosity:
◇Remote transmission distance is greater than 100m need to indicate   
◇Special wetted material mark
◇Perform other flange standard number